ABC of hypertension: Blood pressure measurement : Part I—Sphygmomanometry: factors common to all techniques (1)

This ArticleThe BMJ

BMJ. 2001 Apr 21; 322(7292): 981–985.

PMCID: PMC1120141

PMID: 11312235

ABC of hypertension

Part I—Sphygmomanometry: factors common to all techniques

Gareth Beevers, Gregory Y H Lip, and Eoin O'Brien

Copyright and License information PMC Disclaimer

This article has been corrected. See BMJ. 2001 June 2; 322(7298): 1349.

This article has been corrected. See BMJ. 2001 July 21; 323(7305): 141.

Methods of blood pressure measurement

Most devices for measuring blood pressure are dependent on one common feature, namely, occluding the artery of an extremity (arm, wrist, finger, or leg) with an inflatable cuff to measure blood pressure either oscillometrically, or by detection of Korotkoff sounds. Other techniques, which are not dependent on limb occlusion, such as pulse-waveform analysis, can also be used, but these have little application in clinical practice. The array of techniques available today owe their origins to the conventional technique of auscultatory blood pressure measurement, and these new techniques must indeed be shown to be as accurate as the traditional mercury sphygmomanometer. Since the introduction of sphygmo- manometry, mercury and aneroid sphygmomanometers have been the most popular devices for measuring blood pressures.

This article has been adapted from the newly published 4th edition of ABC of Hypertension. The book is available from the BMJ bookshop and at www.bmjbooks.com

Factors affecting blood pressure measurement

No matter which device is used to measure blood pressure, it must be recognised that blood pressure is a variable haemodynamic phenomenon, which is influenced by many factors, not least being the circ*mstances of measurement itself. These influences on blood pressure can be significant, often accounting for rises in systolic blood pressure greater than 20 mm Hg, and if they are ignored, or unrecognised, hypertension will be diagnosed erroneously and inappropriate management instituted. These factors have to be carefully considered in all circ*mstances of blood pressure measurement—self measurement by patients, conventional measurement, measurement with automated devices whether in a doctor's surgery, an ambulance, a pharmacy, or in hospital using sophisticated technology.1,2

Variability of blood pressure

The observer must be aware of the considerable variability that may occur in blood pressure from moment to moment with respiration, emotion, exercise, meals, tobacco, alcohol, temperature, bladder distension, and pain, and that blood pressure is also influenced by age, race, and circadian variation. It is usually at its lowest during sleep. It is not always possible to modify these many factors but we can minimise their effect by taking them into account in reaching a decision as to the relevance or otherwise of a particular blood pressure measurement.1

Important factors affecting measurement

  • The inherent variability of blood pressure

  • The defence reaction

  • The limitations of the device being used

  • The accuracy of the device

  • Blood pressure is not as easily measured in some groups, such as elderly people

Insofar as is practical the patient should be relaxed in a quiet room at a comfortable temperature and a short period of rest should precede the measurement. When it is not possible to achieve optimum conditions, this should be noted with the blood pressure reading—for example, “BP 154/92, R arm, V phase (patient very nervous).”

“White coat” hypertension

Anxiety raises blood pressure, often by as much as 30 mm Hg. This may be regarded as a physiological reaction, often referred to as the “fight and flight” phenomenon, or “defence” or “alarm” reaction. It is commonly seen in the accident and emergency departments of hospitals when patients are frightened and extremely anxious, but it may also occur in family doctors' surgeries and in the outpatients department. It may occur in normotensive and hypertensive subjects. The degree of this reaction varies greatly from patient to patient, being absent in many, and it is usually reduced or abolished altogether with reassurance and familiarisation with the technique and circ*mstances of blood pressure measurement. Its importance in practice is that decisions to lower blood pressure, and especially to administer drugs, should never be made on the basis of measurements taken in circ*mstances where the defence reaction is likely to be present.

White coat hypertension is a condition in which a normotensive subject becomes hypertensive during blood pressure measurement, but pressures then settle to normal outside the medical environment. It is best demonstrated by ambulatory blood pressure measurement (ABPM).

No one group seems to be exempt from the white coat phenomenon; it may affect the young, the elderly, normotensive and hypertensive subjects, and pregnant women. In young subjects with borderline elevation of conventional blood pressure, identification of white coat hypertension can be of considerable importance in avoiding undue penalties for insurance and employment. Moreover, there are no characteristics that allow for the identification of the phenomenon, other than by obtaining blood pressures away from the medical environment, either by self measurement in the home or with ABPM, which is the technique of choice. Patients diagnosed as “hypertensive” with conventional measurement in whom white coat hypertension is considered a possibility should have ABPM performed before they are labelled “hypertensive,” and certainly before treatment is instigated.

Posture of subject

Posture affects blood pressure, with a general tendency for it to increase from the lying to the sitting or standing position. However, in most people posture is unlikely to lead to significant error in blood pressure measurement provided the arm is supported at heart level. None the less, it is advisable to standardise posture for individual patients and in practice blood pressure is usually measured in the sitting position. Patients should be comfortable whatever their position. No information is available on the optimal time that a subject should remain in a particular position before a measurement, but three minutes is suggested for the lying and sitting positions and one minute standing. Some antihypertensive drugs cause postural hypotension, and when this is expected blood pressure should be measured both lying and standing.1

Arm support

If the arm in which measurement is being made is unsupported, as tends to happen if the subject is sitting or standing, isometric exercise is performed raising blood pressure and heart rate. Diastolic blood pressure may be raised by as much as 10% by having the arm extended and unsupported during blood pressure measurement. The effect of isometric exercise is greater in hypertensive patients and in those taking β blockers. It is essential, therefore that the arm is supported during blood pressure measurement and this is best achieved in practice by having the observer hold the subject's arm at the elbow, although in research the use of an arm support on a stand has much to commend it.1

Arm position

The arm must also be horizontal at the level of the heart as denoted by the midsternal level. Dependency of the arm below heart level leads to an overestimation of systolic and diastolic pressures and raising the arm above heart level leads to underestimation. The magnitude of this error can be as great as 10 mm Hg for systolic and diastolic pressures. This source of error becomes especially important in the sitting and standing positions, when the arm is likely to be dependent by the subject's side. However, it has been demonstrated that even in the supine position an error of 5 mm Hg for diastolic pressure may occur if the arm is not supported at heart level.1,2 Arm position has become an important issue for self measurement of blood pressure with the manufacture of devices for measuring blood pressure at the wrist, which are proving very popular because of the ease of measurement. Many of these devices are inherently inaccurate, but measurement is extremely inaccurate if the wrist is not held at heart level during measurement.

Which arm

This topic remains controversial as some studies, but not all, using simultaneous measurement have demonstrated significant differences between arms.1 However, the fact that blood pressure differences between arms are variable makes the issue even more problematical. A reasonable policy is to measure blood pressure in both arms at the initial examination, and if differences greater than 20 mm Hg for systolic or 10 mm Hg for diastolic pressure are present on three consecutive readings the patient should be referred to a cardiovascular centre for further evaluation.

The cuff and bladder

The cuff is an inelastic cloth that encircles the arm and encloses the inflatable rubber bladder. It is secured around the arm most commonly by means of Velcro on the adjoining surfaces of the cuff, occasionally by wrapping a tapering end into the encircling cuff, and rarely by hooks. Velcro surfaces must be effective, and when they lose their grip the cuff should be discarded. It should be possible to remove the bladder from the cuff so that the latter can be washed from time to time.1

Mismatching of bladder and arm

  • Bladder too small Overestimation of BPUndercuffing

  • Bladder too large Underestimation of BPOvercuffing

  • Undercuffing more common than Overcuffing

“Cuff hypertension”

However sophisticated a blood pressure measuring device may be, if it is dependent on cuff occlusion of the arm (as are the majority of devices), it will then be prone to the inaccuracy induced by miscuffing, whereby a cuff containing a bladder that is either too long or too short relative to arm circumference is used.

A review of the literature on the century-old controversy relating to the error that may be introduced to blood pressure measurement by using a cuff with a bladder of inappropriate dimensions for the arm for which it is intended has shown that miscuffing is a serious source of error, which must inevitably lead to incorrect diagnosis in practice and erroneous conclusions in hypertension research.3 There is unequivocal evidence that either too narrow or too short a bladder (undercuffing) will cause overestimation of blood pressure, so called “cuff hypertension,” and there is growing evidence that too wide or too long a bladder (overcuffing) may cause underestimation of blood pressure. Undercuffing has the effect in clinical practice of overdiagnosing hypertension and overcuffing leads to hypertensive subjects being diagnosed as normotensive. Either eventuality has serious implications for the epidemiology of hypertension and clinical practice.

A proposal for the future—the “Adjustable Cuff”

  • On the basis of a thorough examination of the literature and aware of the advances in cuff design, the design features for an “Adjustable Cuff,” which would be applicable to all adult arms, have been proposed,4 and one such cuff is presently undergoing testing (AC Cosor and Sons Ltd (Surgical), London, UK).

A review of the literature shows that a number of approaches have been used over the years to cope with the difficulty of mismatching and none has been ideal. These have included application of correction factors, a range of cuffs, cuffs containing a variety of bladders, and a cuff for the majority of arms.

Blood pressure measurement in special subjects

Certain groups of people merit special consideration for blood pressure measurement, either because of age, body habitus, or disturbances of blood pressure related to haemodynamic alterations in other parts of the cardiovascular system. Although there is evidence that many subgroups of the hypertensive population may have peculiarities affecting the accuracy of measurement, such as patients with renal disease, patients with diabetes mellitus, women with pre-eclampsia, and youths with “spurious” hypertension, discussion will be confined to children, the elderly, obese subjects, and pregnant women.

Children

Blood pressure measurement in children presents a number of difficulties and variability of blood pressure is greater than in adults, and thus any one reading is less likely to represent the true blood pressure. Also increased variability confers a greater tendency for regression towards the mean. Conventional sphygmomanometry is recommended for general use, but systolic pressure is preferred to diastolic pressure because of greater accuracy and reproducibility. Cuff dimensions are most important and three cuffs with bladders measuring 4×13 cm, 10×18 cm, and the adult dimensions 12×26 cm are required for the range of arm sizes likely to be encountered in the age range 0-14 years. The widest cuff practicable should be used. Korotkoff sounds are not reliably audible in all children under one year and in many under five years of age. In such cases conventional sphygmomanometry is impossible and more sensitive methods of detection such as Doppler, ultrasound, or oscillometry must be used.4

Elderly people

In epidemiological and interventional studies blood pressure predicts morbidity and mortality in elderly people as effectively as in the young.5 The extent to which blood pressure predicts outcome may be influenced by various factors that affect the accuracy of blood pressure measurement and the extent to which casual blood pressure represents the blood pressure load on the heart and circulation.6

The elderly are subject to considerable blood pressure variability, which can lead to a number of circadian blood pressure patterns that are best identified using ambulatory blood pressure measurement. The practical clinical consequence of these variable patterns in the elderly is that blood pressure measuring techniques can be inaccurate and/or misleading.

Pseudohypertension

It has been postulated that as a consequence of the decrease in arterial compliance and arterial stiffening with ageing, indirect sphygmomanometry becomes inaccurate. This has led to the concept of “pseudohypertension” to describe patients with a large discrepancy between cuff and direct blood pressure measurement.7 The significance of this phenomenon has been disputed,8 but in elderly patients in whom blood pressure measured with the conventional technique seems to be out of proportion to the clinical findings, referral to a specialist cardiovascular centre for further investigation may be appropriate.

Overweight people

The association between obesity and hypertension has been known since 1923. The link has been confirmed in many epidemiological studies, and has at least two components.9 Firstly, there appears to be a pathophysiological connection and it may well be that in some cases the two conditions are causally linked, and secondly, if not taken into account, it may result in inaccurate blood pressure values being obtained by indirect measurement techniques.

Obesity may affect the accuracy of blood pressure measurement in children, young people, the elderly, and pregnant women.

The relationship of arm circumference and bladder dimensions has been discussed above. If the bladder is too short, blood pressure will be overestimated—”cuff hypertension”—and, if too long, blood pressure may be underestimated.3

Arrhythmias

The difficulty in measuring blood pressure in patients with arrhythmias is that when cardiac rhythm is irregular there is a large variation in blood pressure from beat to beat. Thus in arrhythmias, such as atrial fibrillation, stroke volume and as a consequence blood pressure vary, depending on the preceding pulse interval. Secondly, in such circ*mstances, there is no generally accepted method of determining auscultatory endpoints. Furthermore blood pressure measuring devices vary greatly in their ability to accurately record blood pressure in patients with atrial fibrillation, indicating that devices should be validated independently in patients with arrhythmias.10

In bradyarrhythmias there may be two sources of error. Firstly, if the rhythm is irregular the same problems as with atrial fibrillation will apply. Secondly, when the heart rate is extremely slow, for example 40 beats per minute, it is important that the deflation rate used is less than for normal heart rates as too rapid deflation will lead to underestimation of systolic and overestimation of diastolic pressure.

Pregnancy

Clinically relevant hypertension occurs in more than 10% of pregnant women in most populations, and in a significant number of these raised blood pressure is a key factor in medical decision making in the pregnancy. Particular attention must be paid to blood pressure measurement in pregnancy because of the important implications for patient management, as well as the fact that it presents some special problems.11

There has been much controversy as to whether the muffling or disappearance of sounds should be taken for diastolic blood pressure. The general consensus from obstetricians based on careful analysis of the evidence is that disappearance of sounds (fifth phase) is the most accurate measurement of diastolic pressure, with the proviso that in those rare instances in which sounds persist to zero the fourth phase of muffling of sounds should be used.12,13

Open in a separate window

Figure

Mercury sphygmomanometer

Open in a separate window

Figure

Example of a normal ambulatory blood pressure pattern plotted by the DABL® Program showing a marked variability of blood pressure

Open in a separate window

Figure

White coat hypertension

Open in a separate window

Figure

Patient in standard seated position

Open in a separate window

Figure

Arm support in standing position

Open in a separate window

Figure

Placement of cuff

Open in a separate window

Figure

Placement of stethoscope

Open in a separate window

Figure

Mean systolic (top) and diastolic (bottom) blood pressures of boys and girls from birth to 18 years. Diastolic blood pressure reflects the use of phase IV Korotkov sounds. Reproduced with permission from de Swiet M, Dillon MJ, Littler W, O'Brien E, Padfield PL, Petrie JC. Measurement of blood pressure in children. Recommendations of a working party of the British Hypertension Society. BMJ 1989;299:469-70

Open in a separate window

Figure

Epidemiological graph for the risk of hypertension in the elderly. Reproduced with permission from Dahloff B, Lindholm LH, Hansson L, Schersten B, Ekbom T, Wester PO. Morbidity and mortality in the Swedish Trial in Old Patients with Hypertension (STOP-Hypertension). Lancet 1991;394:405-12

Open in a separate window

Figure

Recommended bladder length

Open in a separate window

Figure

Blood pressure in atrial fibrillation

Open in a separate window

Figure

Taking blood pressure of a pregnant woman

Table

Recommended bladder dimensions. Data reproduced from O'Brien E, Petrie J, Littler WA et al. Blood Pressure Measurement: Recommendations of the British Hypertension Society. London: BMJ Books, 1997

Dimensions (cm)SubjectMaximum arm circumference (cm)
4×13 Small children17
10×18Medium sized children
Lean adults26
12×26Majority of adult arms33
12×40Obese adults50

Open in a separate window

Accurate readings may be obtained in adults with arm circumferences greater than 50 cm by placing a cuff with a 40 cm bladder so that the centre of the bladder is over the brachial artery. All dimensions have a tolerance of ±1 cm.

References

1. O'Brien E, Petrie J, Littler WA, de Swiet M, Padfield PD, Dillon MJ. Blood pressure measurement: recommendations of the British Hypertension Society. 3rd ed. London: BMJ Publishing Group; 1997. [PMC free article] [PubMed] [Google Scholar]

2. The British Hypertension Society. Blood pressure measurement CD ROM. London: BMJ Books; 1998. [Google Scholar]

3. O'Brien E. A century of confusion: which bladder for accurate blood pressure measurement? J Hum Hypertension. 1996;10:565–572. [PubMed] [Google Scholar]

4. De Swiet M, Dillon MJ, Little W, O'Brien E, Padfield PL, Petrie JC. Measurement of blood pressure in children. Recommendations of a working party of the British Hypertension Society. BMJ. 1989;299:497. [PMC free article] [PubMed] [Google Scholar]

5. Dahloff B, Lindholm LH, Hansson L, Schersten B, Ekbom T, Wester PO. Morbidity and mortality in the Swedish Trial in Old Patients with Hypertension (STOP-Hypertension) Lancet. 1991;394:405–412. [PubMed] [Google Scholar]

6. O'Brien E, O'Malley K. Blood pressure measurement in the elderly with special reference to ambulatory blood pressure measurement. In: Leonetti G, Cuspedi C, editors. Hypertension in the elderly. Dordrecht: Kluwer; 1994. pp. 13–25. [Google Scholar]

7. Spence JD, Sibbald WJ, Cape RD. Pseudohypertension in the elderly. Clin Sci Mol Biol. 1978;55:399s–402s. [PubMed] [Google Scholar]

8. O'Callaghan WG, Fitzgerald D, O'Malley K, O'Brien E. Accuracy of indirect blood pressure measurement in the elderly. BMJ. 1983;286:1545–1546. [PMC free article] [PubMed] [Google Scholar]

9. Staessen J, fa*gard R, Amery A. The relationship between body weight and blood pressure. J Hum Hypertens. 1988;2:207–217. [PubMed] [Google Scholar]

10. Stewart MJ, Gough K, Padfield PL. The accuracy of automated blood pressure measuring devices in patients with controlled atrial fibrillation. J Hypertens. 1995;13:297–300. [PubMed] [Google Scholar]

11. Shennan AH, Halligan AWF. Measuring blood pressure in normal and hypertensive pregnancy. Baillière's Clinical Obstetrics and Gynaecology. 1999;13:1–26. [PubMed] [Google Scholar]

12. Shennan A, Gupta M, Halligan A, Taylor DJ, de Swiet M. Lack of reproducibility in pregnancy of Korotkoff phase IV as measured by sphygmomanometry. Lancet. 1996;347:139–142. [PubMed] [Google Scholar]

13. de Swiet M. K5 rather then K4 for diastolic blood pressure measurement in pregnancy. Hypertension in Pregnancy. 1999;18:iii–v. [PubMed] [Google Scholar]

Articles from The BMJ are provided here courtesy of BMJ Publishing Group

ABC of hypertension: Blood pressure measurement : Part I—Sphygmomanometry: factors common to all techniques (2024)

FAQs

What are the basic techniques of blood pressure measurement? ›

Basic techniques of blood pressure measurement
  • Location of measurement. The standard location for blood pressure measurement is the brachial artery. ...
  • The auscultatory method. ...
  • The oscillometric technique. ...
  • Ultrasound techniques. ...
  • The finger cuff method of Penaz.

What are the factors influencing BP measurement? ›

The observer must be aware of the considerable variability that may occur in blood pressure from moment to moment with respiration, emotion, exercise, meals, tobacco, alcohol, temperature, bladder distension, and pain, and that blood pressure is also influenced by age, race, and circadian variation.

What are some important factors to consider when taking blood pressure measurements? ›

Use this checklist:
  • Don't eat or drink anything 30 minutes before you take your blood pressure.
  • Empty your bladder before your reading.
  • Sit in a comfortable chair with your back supported for at least 5 minutes before your reading.
  • Put both feet flat on the ground and keep your legs uncrossed.
Jan 12, 2024

What are the 3 important steps in measuring blood pressure? ›

How to take blood pressure manually
  • Pump air into the cuff around your arm to tighten it.
  • Slowly release air from the cuff while using a stethoscope to hear the pulse in your arm's brachial artery.
  • Take a measurement when they hear your blood moving.
  • Release more air from the cuff.
  • Measure your pressure again.
Jun 13, 2023

What are three main types of devices used to measure blood pressure describe each one? ›

a cuff that can be inflated with air, a pressure meter (manometer) for measuring air pressure in the cuff, and. a stethoscope for listening to the sound the blood makes as it flows through the brachial artery (the major artery found in your upper arm).

What is the direct method of blood pressure measurement? ›

The technique involves direct measurement of arterial pressure by placing a cannula needle in an artery, usually radial, femoral, or brachial. Two types of probes are generally used. One type is the catheter tip probe and the other is the fluid-filled catheter type.

What 3 factors influence blood pressure and blood flow? ›

Four major factors interact to affect blood pressure: cardiac output, blood volume, peripheral resistance, and viscosity. When these factors increase, blood pressure also increases. Arterial blood pressure is maintained within normal ranges by changes in cardiac output and peripheral resistance.

What are three factors that can affect a patient's blood pressure? ›

The top 10 risk factors for high blood pressure include:
  • Being overweight or obese. The more you weigh the more blood flow you need to supply oxygen and nutrients to your tissues. ...
  • Too much salt (sodium) in your diet. ...
  • Too little potassium in your diet. ...
  • Not being physically active. ...
  • Drinking too much alcohol.

What are the major factors determining blood pressure quizlet? ›

  • What are the 4 factors that affect blood pressure? cardiac output, blood volume, vessel elasticity, and peripheral resistance.
  • Blood volume. It affects blood pressure. ...
  • Vessel Elasticity. ...
  • Peripheral resistance. ...
  • What are the three main sources of peripheral resistance? ...
  • Blood vessel diameter. ...
  • Blood viscosity. ...
  • Hematocrit.

What are three things to avoid before measuring blood pressure? ›

Do not smoke, exercise or use caffeine for 30 minutes to an hour before the test. Such activities increase blood pressure and heart rate. Wear a short-sleeved shirt so that the blood pressure cuff can be placed more easily around your arm. Relax in a chair for at least five minutes before the test.

What are the two most important factors affecting blood pressure? ›

Blood pressure increases with increased cardiac output, peripheral vascular resistance, volume of blood, viscosity of blood and rigidity of vessel walls. Blood pressure decreases with decreased cardiac output, peripheral vascular resistance, volume of blood, viscosity of blood and elasticity of vessel walls.

What is the most important factor in hypertension? ›

Unhealthy Diet

A diet that is too high in sodium and too low in potassium puts you at risk for high blood pressure. Eating too much sodium—an element in table salt—increases blood pressure. Most of the sodium we eat comes from processed and restaurant foods.

How bring down blood pressure quickly? ›

Tricks to Lower Blood Pressure Instantly
  1. Meditate or focus on deep breathing. Meditation and breathing exercises can help you relax, which slows your heart rate and lowers your blood pressure.
  2. Reduce your stress levels. ...
  3. Take a warm bath or shower.

What should you not do before taking blood pressure? ›

You can prepare for the day, but don't eat breakfast or take medications before measuring your blood pressure. If you exercise after waking, take your blood pressure before exercising. Avoid food, caffeine, tobacco and alcohol for 30 minutes before taking a reading. Also, empty your bladder first.

How do you bring down hypertension? ›

10 ways to control high blood pressure without medication
  1. Lose extra pounds and watch your waistline. Blood pressure often increases as weight increases. ...
  2. Exercise regularly. ...
  3. Eat a healthy diet. ...
  4. Reduce salt (sodium) in your diet. ...
  5. Limit alcohol. ...
  6. Quit smoking. ...
  7. Get a good night's sleep. ...
  8. Reduce stress.

What are two methods of measuring blood pressure? ›

The manual method requires auscultation of the blood pressure, whereas the automated system depends on oscillometric devices. With manual blood pressure measurements, both observer and methodological errors can occur.

What is the most effective technique to obtain a blood pressure? ›

The best position to take your blood pressure is seated in a chair with your feet on the floor and your arm supported so your elbow is at about heart level. You should remain in this position throughout the measurement.

Top Articles
Latest Posts
Article information

Author: Eusebia Nader

Last Updated:

Views: 6332

Rating: 5 / 5 (80 voted)

Reviews: 87% of readers found this page helpful

Author information

Name: Eusebia Nader

Birthday: 1994-11-11

Address: Apt. 721 977 Ebert Meadows, Jereville, GA 73618-6603

Phone: +2316203969400

Job: International Farming Consultant

Hobby: Reading, Photography, Shooting, Singing, Magic, Kayaking, Mushroom hunting

Introduction: My name is Eusebia Nader, I am a encouraging, brainy, lively, nice, famous, healthy, clever person who loves writing and wants to share my knowledge and understanding with you.